Centrifuge RPM to RCF Converter – Speed, × g and Rotor k-Factor
Lab protocols say “spin at 500 × g for 5 minutes”, but most benchtop centrifuges are set in revolutions per minute. This centrifuge RPM to RCF converter turns a speed into relative centrifugal force (RCF, in multiples of g) for your rotor, works backwards from a target × g to the RPM to dial in, recovers a rotor radius from the numbers on its label, and converts a protocol's run time from one rotor to another with the k-factor.
Why RCF, not RPM, is the reproducible unit
The force a sample feels depends on both how fast the rotor turns and how far the sample sits from the axis. At 3,000 RPM a tube 5 cm from the axis experiences about 503 × g, while the same speed at 15 cm gives about 1,510 × g. A protocol written in RPM only reproduces when you own the same rotor. RCF describes the force itself, so a pellet formed at 1,000 × g in one lab forms at 1,000 × g in any centrifuge. That is why journals and kit manuals ask for × g.
Deriving the RPM to × g formula
A sample moving in a circle of radius r at angular velocity ω needs a centripetal acceleration of ω²r. Dividing by standard gravity gives the RCF. Speed in RPM converts to radians per second by multiplying by 2π/60:
ω = 2π · RPM / 60 (rad/s)
RCF = r · ω² / g₀ g₀ = 9.80665 m/s²
= 1.11824 × 10⁻⁵ · r(cm) · RPM²
RPM = √( RCF / (1.11824 × 10⁻⁵ · r) )For example, a 10 cm radius at 3,000 RPM gives 1.11824 × 10⁻⁵ × 10 × 3,000² ≈ 1,006 × g. To reach 500 × g with an 8.4 cm rotor you need about 2,307 RPM. The calculator also rounds that to the nearest 10 or 100 RPM dial step and shows the force the rounded speed actually produces. For the underlying physics, see the circular motion calculator; to express the result in other acceleration units, use the acceleration converter.
r_min, r_avg and r_max: which radius?
A tube spans a range of distances from the axis, so the force is not uniform along it. Rotor manuals quote three radii: r_min to the top of the liquid, r_avg to the middle of the column, and r_max to the bottom of the tube. Most protocols and spec sheets state RCF at r_max, so that is the required input here. Adding r_min shows how much weaker the force is at the meniscus. The rotor diagram shades the liquid from light to dark to make that gradient visible, and the RCF-vs-RPM chart draws a band between the r_min and r_max curves.
If you do not know your radius, the Find radius mode solves r = RCF / (1.11824 × 10⁻⁵ · RPM²). Many microcentrifuge lids print a pair such as “13,000 RPM / 15,900 × g”, which gives r_max ≈ 8.41 cm.
The k-factor and pelleting time
Matching the force does not always match the result, because a shorter path from r_min to r_max pellets particles sooner. The k-factor (clearing factor) captures both speed and geometry:
k = 2.533 × 10¹¹ · ln(r_max / r_min) / RPM² (hours)
t = k / S (S in Svedbergs)
t₂ = t₁ · k₂ / k₁A rotor with r_min 4 cm and r_max 8 cm at 50,000 RPM has k ≈ 70.2, so a 70 S ribosome pellets in about one hour. If your rotor has k = 35.1, a 60-minute protocol needs only 30 minutes. The rotor to rotor mode does this for you, finds the RPM that matches the original RCF at r_max, and optionally gives pelleting times for a sedimentation coefficient you enter. Lower k means faster clearing; the relation assumes sedimentation in a medium like water and is a guide, not a guarantee, for dense gradients.
Reading the application gauge
The log-scale gauge places your setting against typical ranges: a few hundred × g for gentle cell pelleting, several thousand for bacteria and yeast, 10,000–16,000 × g in a microcentrifuge for precipitated DNA, and 100,000 × g and above for membranes and ribosomes in an ultracentrifuge. These bands are approximate and vary between sources, so always follow your protocol. When preparing the samples themselves, the dilution calculator and buffer preparation calculator cover the bench maths before the spin.
Sharing and saving
Every input is stored in the page address, so Copy link shares an exact protocol conversion with a colleague. Rotors you save (name, r_min, r_max and rated speed) stay in your browser only. The quick reference table lists RCF at common speeds and RPM for common forces and copies as tab-separated values for a lab notebook or spreadsheet.