Logo

MonoCalc

/

Centrifuge RPM to RCF Converter

Chemistry
Unit for every radius below

Rotor radius

Most protocols quote RCF at r_max.
Axis to the top of the liquid.
Leave blank to use (r_min + r_max) / 2.
Used to flag speeds above the rating.

Read r_min and r_max from your rotor manual or lid label; radii depend on the rotor model and any tube adapter. Saved rotors stay in this browser only.

RCF at r_max (9.5 cm)

956 × g

9,376 m/s²

Angular velocity ω

314.2 rad/s

ω = 2π · RPM / 60

RCF = r·ω²/g₀ with ω = 2π·RPM/60 and g₀ = 9.80665 m/s², i.e. 1.11824 × 10⁻⁵ · r(cm) · RPM², computed from the exact expression. References that round the constant to 1.118 × 10⁻⁵ or 1.12 × 10⁻⁵ can differ in the last digit.

Safety
Never exceed the rotor's rated maximum speed, and derate it for dense samples as the manufacturer specifies.
axisr_max 9.5 cm · 956 × g

Liquid level is illustrative: add r_min to place it. Shading: lighter near the top of the liquid (lower g), darker at the tube bottom (higher g).

RCF vs RPM

RCF rises with the square of speed: doubling RPM gives four times the force.

Where 956 × g sits

Typical ranges — always follow your protocol

956 × g

10²

10³

10⁴

10⁵

10⁶

Gentle mammalian cell pelleting: 100–500 × g

Removing whole cells / nuclei from homogenate: 500–1,000 × g

Bacteria / yeast pelleting: 3,000–6,000 × g

Pelleting precipitated DNA/protein (microcentrifuge): 10,000–16,000 × g

Mitochondria / organelles: 10,000–20,000 × g

Microsomes, membranes: 80,000–120,000 × g

Ribosomes, macromolecules (ultracentrifuge): 100,000–600,000 × g

Log scale, 100 to 1,000,000 × g. Ranges are approximate and differ between protocols and sources.

Quick reference at r_max = 9.5 cm

RPMRCF (× g)
1,000106
2,000425
3,000956
4,0001,700
5,0002,656
10,00010,623
13,00017,953
15,00023,902
RCF (× g)RPM
100970
3001,680
5002,169
1,0003,068
3,0005,314
10,0009,702

About This Tool

Centrifuge RPM to RCF Converter – Speed, × g and Rotor k-Factor

Lab protocols say “spin at 500 × g for 5 minutes”, but most benchtop centrifuges are set in revolutions per minute. This centrifuge RPM to RCF converter turns a speed into relative centrifugal force (RCF, in multiples of g) for your rotor, works backwards from a target × g to the RPM to dial in, recovers a rotor radius from the numbers on its label, and converts a protocol's run time from one rotor to another with the k-factor.

Why RCF, not RPM, is the reproducible unit

The force a sample feels depends on both how fast the rotor turns and how far the sample sits from the axis. At 3,000 RPM a tube 5 cm from the axis experiences about 503 × g, while the same speed at 15 cm gives about 1,510 × g. A protocol written in RPM only reproduces when you own the same rotor. RCF describes the force itself, so a pellet formed at 1,000 × g in one lab forms at 1,000 × g in any centrifuge. That is why journals and kit manuals ask for × g.

Deriving the RPM to × g formula

A sample moving in a circle of radius r at angular velocity ω needs a centripetal acceleration of ω²r. Dividing by standard gravity gives the RCF. Speed in RPM converts to radians per second by multiplying by 2π/60:

ω = 2π · RPM / 60 (rad/s) RCF = r · ω² / g₀ g₀ = 9.80665 m/s² = 1.11824 × 10⁻⁵ · r(cm) · RPM² RPM = √( RCF / (1.11824 × 10⁻⁵ · r) )

For example, a 10 cm radius at 3,000 RPM gives 1.11824 × 10⁻⁵ × 10 × 3,000² ≈ 1,006 × g. To reach 500 × g with an 8.4 cm rotor you need about 2,307 RPM. The calculator also rounds that to the nearest 10 or 100 RPM dial step and shows the force the rounded speed actually produces. For the underlying physics, see the circular motion calculator; to express the result in other acceleration units, use the acceleration converter.

r_min, r_avg and r_max: which radius?

A tube spans a range of distances from the axis, so the force is not uniform along it. Rotor manuals quote three radii: r_min to the top of the liquid, r_avg to the middle of the column, and r_max to the bottom of the tube. Most protocols and spec sheets state RCF at r_max, so that is the required input here. Adding r_min shows how much weaker the force is at the meniscus. The rotor diagram shades the liquid from light to dark to make that gradient visible, and the RCF-vs-RPM chart draws a band between the r_min and r_max curves.

If you do not know your radius, the Find radius mode solves r = RCF / (1.11824 × 10⁻⁵ · RPM²). Many microcentrifuge lids print a pair such as “13,000 RPM / 15,900 × g”, which gives r_max ≈ 8.41 cm.

The k-factor and pelleting time

Matching the force does not always match the result, because a shorter path from r_min to r_max pellets particles sooner. The k-factor (clearing factor) captures both speed and geometry:

k = 2.533 × 10¹¹ · ln(r_max / r_min) / RPM² (hours) t = k / S (S in Svedbergs) t₂ = t₁ · k₂ / k₁

A rotor with r_min 4 cm and r_max 8 cm at 50,000 RPM has k ≈ 70.2, so a 70 S ribosome pellets in about one hour. If your rotor has k = 35.1, a 60-minute protocol needs only 30 minutes. The rotor to rotor mode does this for you, finds the RPM that matches the original RCF at r_max, and optionally gives pelleting times for a sedimentation coefficient you enter. Lower k means faster clearing; the relation assumes sedimentation in a medium like water and is a guide, not a guarantee, for dense gradients.

Reading the application gauge

The log-scale gauge places your setting against typical ranges: a few hundred × g for gentle cell pelleting, several thousand for bacteria and yeast, 10,000–16,000 × g in a microcentrifuge for precipitated DNA, and 100,000 × g and above for membranes and ribosomes in an ultracentrifuge. These bands are approximate and vary between sources, so always follow your protocol. When preparing the samples themselves, the dilution calculator and buffer preparation calculator cover the bench maths before the spin.

Rotor safety
Never run a rotor above its rated maximum speed. Rotors must be derated for samples denser than the manufacturer's reference density, and for older or damaged rotors. Save each rotor with its rated speed and the tool flags any conversion that would exceed it.

Sharing and saving

Every input is stored in the page address, so Copy link shares an exact protocol conversion with a colleague. Rotors you save (name, r_min, r_max and rated speed) stay in your browser only. The quick reference table lists RCF at common speeds and RPM for common forces and copies as tab-separated values for a lab notebook or spreadsheet.

Frequently Asked Questions

Is the Centrifuge RPM to RCF Converter free?

Yes, Centrifuge RPM to RCF Converter is totally free :)

Can I use the Centrifuge RPM to RCF Converter offline?

Yes, you can install the webapp as PWA.

Is it safe to use Centrifuge RPM to RCF Converter?

Yes, any data related to Centrifuge RPM to RCF Converter only stored in your browser (if storage required). You can simply clear browser cache to clear all the stored data. We do not store any data on server.

How does this centrifuge RPM to RCF converter work?

It uses RCF = r·ω²/g₀ with ω = 2π·RPM/60 and g₀ = 9.80665 m/s², which for a radius in centimetres is RCF ≈ 1.11824 × 10⁻⁵ · r · RPM². The same equation is rearranged to find the RPM for a target RCF or the radius from a rotor's rated RPM and RCF.

Why do protocols give × g instead of RPM?

The force on a sample depends on both speed and rotor radius, so the same RPM gives very different forces in a microcentrifuge and a large floor centrifuge. RCF in × g describes the force itself, so any lab can reproduce the spin whatever rotor it has.

Which radius should I enter: r_min, r_avg or r_max?

Enter r_max, the distance from the rotor axis to the bottom of the tube, because most protocols and rotor spec sheets quote RCF there. Add r_min (axis to the top of the liquid) to see how much lower the force is at the top of the tube and to calculate the k-factor.

Where do I find my rotor's radius?

Rotor manuals and lid labels list r_min, r_avg and r_max, usually in millimetres. If you only know the rotor's maximum RPM and the RCF it gives, use the Find radius mode to work out r_max from those two numbers.

What is a rotor k-factor?

The k-factor, or clearing factor, measures how quickly a rotor pellets particles at a given speed: k = 2.533 × 10¹¹ · ln(r_max/r_min) / RPM². A particle with sedimentation coefficient S pellets in about k/S hours, and a run can be moved to another rotor with t₂ = t₁ · k₂/k₁.

Why does my result differ slightly from another calculator?

Many references round the constant to 1.118 × 10⁻⁵ or 1.12 × 10⁻⁵, which shifts the last digit. This tool computes from the exact expression, and the k-factor time conversion assumes sedimentation through the full path from r_min to r_max in a fluid like water.