Chromatography Resolution Calculator

Chemistry

Your data never leaves this browser.

Method settings

min
The first baseline disturbance, or an unretained marker (uracil or thiourea in reversed phase)
Optional, for H and N/m
µm
For the reduced plate height h

Peaks

Namet_R (min)W₀.₅ (min)HeightTailing T
Remove

Height is relative and optional; T is the USP tailing factor at 5 % height.

Critical pair: API – Imp B

Resolution Rs

1.48

Partially resolved · below 2.00

Selectivity α

1.074

Purnell Rs_P

1.45

√N/4 × (α−1)/α × k/(1+k)

Valley (% of Imp B)

12.6 %

p/v 7.93

32.8 × 0.0685 × 0.646 = 1.45

efficiency × selectivity × retention. Purnell assumes both peaks share the later peak's width and are Gaussian, so it differs slightly from the measured Rs.

Crossover: API 0.07 % into Imp B; Imp B 0.77 % into API.

Net perpendicular-drop error: API −0.01 %, Imp B +0.10 %.

Modelled from Gaussian / bi-Gaussian shapes built from your widths, heights and tailing. Real peaks (exponential tails, baseline drift) differ, so use these figures to judge a separation, not to report one.

Rs 0.5

15.9 % crossover

Rs 1.0

2.3 % crossover

Rs 1.5

0.13 % crossover

Rs 2.0

< 0.01 % crossover

Simulated chromatogram

Hatched: the critical pair's crossover. Modelled from Gaussian / bi-Gaussian shapes built from your widths, heights and tailing. Real peaks (exponential tails, baseline drift) differ, so use these figures to judge a separation, not to report one.

Peak results

Peakt_R (min)kNN/mH (µm)hN_effTFlags
Imp A3.851.4115,840105,6039.472.715,4101.10
API4.321.7016,994113,2918.832.526,7371.30
Imp B4.521.8317,251115,0078.702.487,2001.20
Imp C7.13.4420,057133,7127.482.1412,0361.00

N from 5.54 (t_R / W₀.₅)²; N_eff uses t_R − t₀. Flags: fronting for T < 0.8, tailing for T > 1.5.

Adjacent pairs

PairΔt_R (min)αRsBandValleyCrossover a / bNet drop error a / b
Imp A – API0.471.2093.70RobustBaseline< 0.01 % / < 0.01 %0.00 % / 0.00 %
API – Imp Bcritical
0.21.0741.48Partially resolved12.6 %0.07 % / 0.77 %−0.01 % / +0.10 %
Imp B – Imp C2.581.88415.30RobustBaseline< 0.01 % / < 0.01 %0.00 % / 0.00 %

What it would take: API – Imp B to Rs 2.00

Required Rs comes from the system-suitability threshold below. Each curve is the Purnell equation scaled to pass through your measured Rs, with the other two levers held.

LeverTo reach Rs 2.00
PlatesN ≈ 31,300 (×1.82): a 272 mm column, run time ×1.82 (same flow and particle size), or the same 150 mm column with 1.93 µm particles; back-pressure ×3.30
Selectivityα ≈ 1.102 (now 1.074)
Retentionk ≈ 6.72 (now 1.83); the later peak would elute at ≈ 12.4 min

Plates

Selectivity

Retention

Raising α by 0.01 does as much as multiplying N by ×1.27.

System suitability

Screening check. Your method's own system-suitability criteria take precedence.

Required Rs:

Rs ≥ 2.00

Fail

API – Imp B: Rs 1.48 (critical pair)

k ≥ 2.00

Fail

Below: Imp A, API, Imp B

N ≥ 2,000

Pass

All 4 checked peaks pass

T ≤ 2.00

Pass

4 peaks with T entered pass

Replicate-injection %RSD: use the Coefficient of Variation Calculator.

About This Tool

Chromatography Resolution Calculator – Read a Chromatogram and Know What to Change

Every HPLC or GC method lives or dies by its critical pair, the two neighbouring peaks that are hardest to pull apart. This chromatography resolution calculator takes the peak table you already have (retention time and width for each peak, plus the dead time) and returns the resolution Rs, plate count N, retention factor k and selectivity α for every adjacent pair. It draws the chromatogram those numbers describe and shows how far each of the three levers would have to move to reach Rs 1.5 or 2.0. Everything is calculated in your browser.

What Rs measures

Resolution compares the gap between two apexes with the width of the peaks. With half-height widths, as in the harmonised USP <621> and Ph. Eur. 2.2.46 text, it is:

Rs = 1.18 · (t_R,b − t_R,a) / (W₀.₅,a + W₀.₅,b)

With baseline tangent widths the constant is 2 instead. The rule of thumb is simple: below Rs 1.0 the peaks overlap noticeably, 1.5 is baseline resolution for equal Gaussian peaks, and 2.0 leaves a margin for tailing, column ageing and unequal peak sizes. The reference thumbnails in the tool show what each level looks like, with the share of each peak that crosses into its neighbour: about 2.3 % at Rs 1.0 and 0.13 % at Rs 1.5.

Half-height or tangent widths, and why data systems disagree

The plate count is N = 5.54 (t_R / W₀.₅)² from the half-height width, or N = 16 (t_R / W_b)² from the tangent width. For a perfect Gaussian both give the same answer. Real peaks tail, and then the method matters: tangent widths, the 5σ method at 4.4 % height and statistical moments all weigh the tail more heavily than the half-height width does, so they report lower plate counts. Pick the width definition you actually measured, and compare plate counts only between reports that use the same one.

Retention factor and dead time

The retention factor k = (t_R − t₀) / t₀ tells you how much longer a compound stays on the column than the mobile phase does. You need the dead time t₀: the first baseline disturbance, an unretained marker such as uracil in reversed-phase LC, or methane in GC. If you have none, the calculator estimates t₀ from the column diameter, length, flow rate and total porosity.

The Purnell equation, term by term

Rs = (√N / 4) · ((α − 1) / α) · (k / (1 + k))

The three factors are efficiency, selectivity and retention, using N and k of the later peak. The tool shows them multiplied out for your critical pair, for example 32.8 × 0.0685 × 0.646 = 1.45, next to the measured Rs. The two differ slightly because Purnell assumes both peaks share the later peak's width.

Why selectivity beats plates

Plates enter under a square root, so doubling the column length buys only 41 % more resolution, at twice the run time and pressure. Selectivity works far harder: at α = 1.074, raising α by just 0.01 does as much as multiplying N by about 1.27. The calculator prints that equivalence for your own pair. In LC the practical selectivity levers are the organic modifier, the pH near the analytes' pKa, the stationary-phase chemistry and the temperature; in GC they are the phase polarity and the temperature programme.

Why more retention buys little above k ≈ 10

The retention term k / (1 + k) is 0.50 at k = 1, 0.83 at k = 5 and 0.91 at k = 10, then creeps towards 1. Pushing k from 5 to 10 raises Rs by only 9 % but nearly doubles the run time. When the required k comes out impossibly high, the tool says so and gives the ceiling Rs that infinite retention would reach.

Gradient elution and peak capacity

In gradient or temperature-programmed runs peaks are compressed as they travel, so N and k from t_R and W lose their meaning. Resolution from widths is still valid, and the calculator adds the peak capacity n_c = 1 + t_G / W̄_b, roughly the largest number of peaks the gradient could separate at Rs 1.

TLC's R_F is not the retention factor k
Thin-layer chromatography reports a retardation factor R_F, the distance a spot travels divided by the solvent front distance. It runs from 0 to 1 and is a different quantity from the column retention factor k used here, although both describe how strongly a compound is held by the stationary phase.

Frequently Asked Questions

Is the Chromatography Resolution Calculator free?

Yes, Chromatography Resolution Calculator is totally free :)

Can I use the Chromatography Resolution Calculator offline?

Yes, you can install the webapp as PWA.

Is it safe to use Chromatography Resolution Calculator?

Yes, any data related to Chromatography Resolution Calculator only stored in your browser (if storage required). You can simply clear browser cache to clear all the stored data. We do not store any data on server.

What is chromatographic resolution, and what Rs counts as baseline separation?

Resolution Rs is the distance between two peak apexes divided by their average width, so it measures separation in units of peak width. For two equal Gaussian peaks, Rs 1.0 leaves about 2.3 % of each peak overlapping the other, Rs 1.5 is conventionally called baseline resolution (about 0.13 % overlap), and Rs 2.0 is the usual target for a robust method.

Should I measure peak width at half height or at the baseline?

Half height is easier to measure reliably, because it sits well clear of baseline noise and drift, and it is what the harmonised USP <621> and Ph. Eur. 2.2.46 formulas use (N = 5.54 (tR/W₀.₅)², Rs = 1.18 ΔtR/(W₀.₅,₁ + W₀.₅,₂)). Baseline tangent widths use the constants 16 and 2 instead. Either works here, as long as you select the definition you actually measured.

Why doesn't my plate count match my data system's?

Data systems differ in how they measure width: tangent, half height, 4.4 % height (the 5σ method, N = 25 (tR/W)²) or statistical moments. For a perfect Gaussian they agree, but any tailing makes the low-height and moment methods give noticeably smaller N than the half-height method. Compare like with like, and check which method your report uses.

What is the retention factor k, and how do I find the dead time t₀?

The retention factor k = (tR − t₀)/t₀ is how much longer a compound spends on the column than an unretained one, so it is independent of flow rate and column length. Read t₀ from the first baseline disturbance or inject an unretained marker such as uracil or thiourea in reversed-phase LC, or methane or air in GC. The calculator can also estimate t₀ from the column dimensions and flow.

What is the best way to improve resolution: more plates, more retention or better selectivity?

Usually selectivity. Rs grows only with the square root of N, so doubling the column length buys about 41 % more resolution at twice the run time, and retention gains flatten out above k ≈ 5–10. A small change in α, from a different organic modifier, pH, stationary phase or temperature, is often worth more plates than any practical column.

Can I use these formulas for gradient or temperature-programmed runs?

Resolution measured from retention times and widths is still valid, so the calculator keeps Rs, the valley analysis and the tailing factor. Plate counts and retention factors from tR and W are not meaningful there, because peaks are compressed as they move, so those outputs are hidden; peak capacity from the gradient time takes their place.