Kjeldahl Nitrogen Calculator – From Titration to % Nitrogen and % Protein
The Kjeldahl method is the reference technique for total organic nitrogen and, from it, crude protein in food, feed, soil, fertiliser and wastewater laboratories. This Kjeldahl nitrogen calculator turns your sample mass, burette readings and titrant concentration into %N and % protein, and shows every step so you can compare the result with your lab notebook.
How the Kjeldahl method works
A weighed sample is digested in hot sulfuric acid with a catalyst, converting organic nitrogen into ammonium sulfate. The digest is made alkaline with NaOH and the liberated ammonia is distilled into a receiving solution. The ammonia is then measured by titration in one of two ways.
- Boric acid (direct) titration: NH₃ is trapped in boric acid and titrated directly with standard acid.
- Back titration: NH₃ is trapped in a known excess of standard acid and the unreacted acid is titrated with standard NaOH.
Formulas used
With volumes in mL and mass in g, mmol × 14.0067 gives mg of nitrogen:
%N = (V_s − V_b) × N_acid × 1.40067 / W for the direct method.
mmol NH₃ = V_acid × N_acid − V_NaOH × N_NaOH for a back titration without a blank, or (V_blank − V_sample) × N_NaOH when a blank is run.
% protein = %N × F, and the dry-basis value divides by (1 − moisture / 100).
Why the protein factor matters
Kjeldahl measures nitrogen, not protein. Protein is estimated by multiplying %N by a factor that assumes a typical nitrogen content of the protein. The traditional 6.25 assumes 16 % nitrogen, but real proteins differ. Jones factors, as tabulated by FAO, use 6.38 for dairy, 5.70 for refined wheat flour, 5.71 for soybeans and 5.18 for almonds. Mariotti et al. (2008) argue that 6.25 overestimates protein in most foods and suggest 5.6 as a general value. Always state the factor you used when reporting protein.
Blanks, replicates and sanity checks
A reagent blank corrects for nitrogen in the reagents and for titrant used to reach the endpoint. Running replicates lets you see the mean, standard deviation and relative standard deviation; an RSD above about 2 % suggests poor agreement. In a back titration, if the ammonia consumes more than about 90 % of the acid in the receiver, some ammonia may have escaped, so repeat with more acid or less sample.
Limitations
Standard Kjeldahl does not recover nitrate or nitrite nitrogen; modified procedures such as Devarda or salicylic acid digestion are needed. Non-protein nitrogen, such as melamine, inflates apparent protein. Combustion (Dumas) analysis is an alternative technique with its own conversion factors. This tool covers only the arithmetic after titration.